7 - 44
Section 7-5 - Gathering Troubleshooting Data (R4.2.5x and later)
7-5-5
Capturing network logs to use for troubleshooting
Use the
Network Capture
utility to capture network logs. Once the logs are captured and extracted from
the system, use Microsoft Message Analyzer to convert the output, and then Wireshark (or some other
network sniffer) to analyze it.
To troubleshoot network capture logs
:
1.) On the LOGIQ™ S8, run the
Network Capture
utility to generate .etl and .cab files for analysis. For
more information, see
"Network Capture" on page 7-61
.
2.) Locate the .etl file at D:\Service.
3.) Transfer these files to a laptop.
4.) On the laptop, download Microsoft Message Analyzer (if not already done).
5.) In Microsoft Message Analyzer, open the .etl file.
If the DICOM connection is encrypted, you will not be able to see the contents of the packets.
Microsoft Message Analyzer does not have a plug-in to decode DICOM packets.
Содержание 5418099
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO ...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x ...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank ...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16 ...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork ...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4 ...
Страница 417: ......