Section 7-3 - Gathering Troubleshooting Data (R1 to R4)
7 - 17
7-3-7 Capturing Network Logs with Network Sniffer (cont’d)
10.) Press Alt+N to restore Sniffer window and observe the network activity.
Figure 7-16 Sniffer Window and Network Activity
A.) “Packet List” pane - the packet list pane displays all the packets in the current capture file. Each line
in the packet list corresponds to one packet in the capture file.
B.) “Packet Details” pane - shows the current packet (selected in the “Packet List” pane) in a more
detailed form. This pane shows the protocols and protocol fields of the packet selected in the
“Packet List” pane.
C.) “Packet Bytes” pane - The packet bytes pane shows the data of the current packet (selected in the
“Packet List” pane) in a hexdump style.