Section 1-4 - Safety Considerations
1-4-7
Battery Safety
NOTE:
The LOGIQ S8 ultrasound system is supplied with a lithium ion battery in the power supply
module, as option.
The lithium ion technology used in the system’s battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries (such as watch batteries).
Used batteries should not be placed with common household waste products. Contact local authorities
for the location of a chemical waste collection program nearest you.
NOTE:
Regulations vary for different countries. Dispose of a used battery in accordance with local
regulations.
•
DO NOT short-circuit the battery by directly connecting the negative terminals with metal objects.
•
DO NOT heat the battery or discard it in a fire.
•
DO NOT expose the battery to temperatures over 50 °C. Keep the battery away from fire and other
heat sources.
•
DO NOT charge the battery near a heat source, such as, a fire or heater.
•
DO NOT leave the battery in direct sunlight.
•
DO NOT pierce the battery with a sharp object, hit it, or step on it.
•
DO NOT use a damaged battery.
•
DO NOT apply solder to a battery.
•
DO NOT connect the battery to an electrical power outlet.
•
DO NOT immerse the battery in water or allow it to get wet.
•
DO NOT place the battery into a microwave oven or pressurized container.
•
If the battery leaks or emits an odor, remove it from all possible flammable sources.
•
If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during
use, recharging or storage, immediately remove it and stop using it.
NOTE:
If you have any questions about the battery, consult your local GE representative.
•
Storage of the battery pack:
Short-term (less than 1 month): between 0 °C (32 °F) and 50 °C (122 °F)
Long-term (more than 3 months): between 10 °C (50 °F) and 35 °C (95 °F)
When charging the battery for the first time after long-term storage, recover the battery pack to original
performance through repeating several cycles of full charging and discharging.
WARNING
WARNING
THE LOGIQ S8 BATTERY HAS A SAFETY DEVICE.
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY!
ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS:
CAUTION
USE ONLY BATTERIES APPROVED BY GE HEALTHCARE AS SUITABLE FOR USE WITH THE
LOGIQ S8/LOGIQ E8 ULTRASOUND SYSTEM.
CAUTION
TO PREVENT THE BATTERY BURSTING, IGNITING, OR FUMES FROM THE BATTERY CAUSING
EQUIPMENT DAMAGE, ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS:
Содержание 5418099
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO ...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x ...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank ...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16 ...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork ...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4 ...
Страница 417: ......