GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 10 Care & Maintenance
10-7
10-5-2-2
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted. Refer
to the User Manual for a list of approved peripherals/options.
10-5-3
Input Power
10-5-3-1
AC/DC Adapter Inspection
10-5-4
Cleaning
10-5-4-1
General Cleaning
Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks
Step
Item
Description
1
B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
2
Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
3
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
4
Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
5
ECG
Verify basic operation with customer.
6
DVD
Verify that the DVD is functioning properly. Clean heads and covers if necessary.
Table 10-7 AC/DC Adapter Inspection
Step
Item
Description
1
Unplug Cord Disconnect the mains cable from the wall and system.
2
Inspect
Inspect it and its connectors for damage of any kinds.
3
Verify
Verify that the LINE wires are properly attached to the terminals, and that no strands may cause a short
circuit.
Table 10-8 General Cleaning
Step
Item
Description
1
Console
Remove the battery. Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire
system. Be careful not to get the cloth too wet so that moisture does not enter the console.
2
Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
3
LCD
Use a soft, non-abrasive folder cloth. Gently wipe the LCD face. DO NOT use a glass cleaner that has a
hydrocarbon base (such as Benzene, Methy Alcohol or Methy Ethyl Ketone) on LCD with the filter (anti-
glare shield).