GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
2-2
Section 2-2 - General Console Requirements
Section 2-2
General Console Requirements
2-2-1
Console Environmental Requirements
2-2-1-1
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and diameters can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interface.
2-2-2
Electrical Requirements
NOTE:
GE Medical Systems requires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within the facility, the ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-2-1
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Power Requirements
Table 2-2
Environmental Requirements for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Scanners
Operational
Storage
Transport
Temperature
10 - 40 degree C
-5 - 50 degree C
-5 - 50 degree C
Humidity
30 - 75%
non-condensing
10 - 90%
non-condensing
10 - 90%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Temperatures in degree C, conversion to degree F =(degree C*(9/5) + 32)
Table 2-3
Electrical Specifications for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
Adapter
Voltage
Power
Current
Frequency
TWADP100
100-240 VAC
130VA max.
1.3 A (max.)
50/ 60HZ