Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 5 Components and Functions (Theory)
5-11
Every access request, whether successful or not, will be logged into a service access log that is
viewable to authorized users.
5-4-2-4
The usage for security cable
The ultrasound system equipped with Kensington security slot which is compatible with a Kensington
security cable, refer to
How to prevent unauthorized removal of the ultrasound system?
1.) Wrap the cable around the immovable object, refer to
;
2.) Make sure and rotate the key to the right (unlocked position);
3.) Insert the lock into the Kensington security slot in the system side cover, refer to
4.) Rotate the key to the left (locked position).
5.) For more information, visit www.kensington.com.
NOTE:
Full maintenance reboot should be selected after the service dongle is used on the system.
Table 5-2 Service Login User Levels
User Level
Access Authorization
Password
Operator
Authorized access to specified diagnostics, error logs and utilities. Same
acquisition diagnostic tests as GE Service.
uls
Administrator
uls
External Service
gogems
Figure 5-8 Kensington security slot
Figure 5-9 Kensington security slot and system