SERVICE MANUAL
Medi-Therm II
w
w
w
.g
aymar
.com P/N 10187-000 7/98
TROUBLESHOOTING CHARTS
**
Is a steel object
attracted to center top
of the valve?
*
*
Yes
Replace
impeller.
No
Is
impeller
broken or
cracked
?
Yes
No
No
Yes
Repair/replace
hot solenoid SV2.
Suspect
Q8 or U7.
Repair/replace
power supply
board.
Assume
Q8 triac
is also damaged.
Repair/replace
power supply
board.
Is
hot solenoid
SV2 coil
shorted
?
Disconnect
hot solenoid SV2.
Measure coil
resistance for short.
Is
cold solenoid
SV1 coil
shorted
?
Suspect
Q9 or U8.
Repair/replace
power supply
Assume
Q9 triac is
also damaged.
Repair/replace
power supply board.
Repair/replace
cold solenoid
SV1.
Yes
Yes
No
No
No
Yes
Yes
No
Yes
Call
GAYMAR
Technical
Service
No
Repair/replace cold
solenoid SV1.
Suspect
control/display
board.
Repair/replace.
Is SV1 coil
energized?
Is pin 9
ref to pin 10
of U51
0 VDC?
Suspect
board-to-board
connection.
Repair.
Is
optocoupler
pulled
low?
Line
voltage at
output of
Q8?
Suspect
base-to-power
supply board
connection.
Line
voltage at
terminals 2
and 3 of
TB1?
U7 input
Figure 6I—Check Water Flow Alert is On, Page 2 of 2
49