SERVICE MANUAL
Medi-Therm II
w
w
w
.g
aymar
.com P/N 10187-000 7/98
No
Ye
s
Ye
s
No
No
Is S1
thermostat
out of spec
?
T
est S1
as indicated in
Low T
emperature
Thermostat
Function T
est.
Ye
s
No
Ye
s
No
No
Ye
s
Ye
s
Ye
s
No
Ye
s
No
Replace
thermostat.
Replace S1
Does
TPT9 read
less than 37.4°F
(3°C) ?
Is
S2 or S3
thermostat out
of spec
?
Does
RFU condition
reappear?
Does
TPT9 read
less than 37.4°F
(3°C)?
T
est S2, S3 as
indicated in High
T
e
mperature
Thermostat
Function T
est.
Operate unit in
MANUAL mode
with setpoint
temperature equal
to 39.2°F (4°C).
Does
RFU condition
reappear?
Possible faulty control system
or temperature measurement
circuitry
. Call Gaymar
T
echnical Services.
Allow unit to cool
to room
temperature or
until
thermostat closes.
Allow unit to warm
to room
temperature or
until thermostat
closes.
Operate unit in
MANUAL mode
with setpoint
temperature equal
to 39.2°F (4°C).
Does
RFU condition
reappear?
Is it S2 or S3?
S1 is open
TROUBLESHOOTING CHARTS
Figure 6H—RFU Code H, Page 2 of 2
47