Figure 6E—RFU Codes 6, 7
TROUBLESHOOTING CHARTS
42
RFU CODE 7
Compensation resistor 2
(R13 on control/display
board) reading is too low.
No
Yes
Repair/replace
the
control/display board.
Suspect
temperature
measurement
circuitry.
Is it
7320 ohms
±0.1%
?
Replace R13.
Measure R13
resistance.
RFU CODE 6
Compensation resistor 2
(R13 on
control/display board)
reading is too high.
SERVICE MANUAL