SERVICE MANUAL
Medi-Therm II
REPAIR PROCEDURES
w
w
w
.g
aymar
.com P/N 10187-000 7/98
Always perform the
FUNCTIONAL CHECK AND SAFETY INSPECTION
(section 7.3) after making repairs and before returning the Medi-Therm
II machine to patient use.
Improper repair may result in death or serious injury,
equipment damage, or malfunction.
In addition to the following repair procedures, refer to the troubleshooting charts
in section 8.0, page 34,
TROUBLESHOOTING AND SERVICE MODES.
•
Equipment contains refrigerant under high pressure. Refer
servicing to a certified refrigeration service technician.
Improper repair procedures can result in serious
injury or damage the compressor unit.
•
The refrigeration system contains rotating fan blades. Do not
operate with fan shroud removed.
Injury could result.
NOTE: Servicing and/or repair of the refrigeration system should be
performed in compliance with applicable federal, state, and local
regulations.
The Medi-Therm II refrigeration system uses either an R-12 (CFC-12) or an
R-134a (HFC-134a) refrigerant, but not both. To determine which type of
refrigerant is required by your system, refer to the nameplate located on the base
of the compressor. If you are still unable to identify the refrigerant type, contact
Gaymar’s Technical Service Department for assistance.
Refer to the appropriate section (9.1.1 or 9.1.2, p. 58) for charging
specifications.
9.0
REPAIR PROCEDURES
WARNING
WARNING
9.1
REFRIGERATION
SYSTEM
57