.com P/N 10187-000 7/98
Figure 6G—RFU Code 9
TROUBLESHOOTING CHARTS
44
Yes
No
Faulty signal lines
between
control/display
board and RT1.
Repair/replace
the
control/display board.
Repair/replace
the
control/display
board.
Suspect
temperature
measurement
circuitry.
Is it
between
811 and 7355
ohms
?
Run machine. If RFU condition
(RFU Code 9) reappears, unplug
machine and measure resistance
between pins 23 and 11 of J1 on
back of control/display board.
Insure integrity of control/display
board to power supply board
connections. Insure integrity of
base to power supply board
connections.
Yes
Yes
No
No
Suspect controller failure
caused RT1 to actually
read outside
32°F to 122°F
(0°C to 50°C) range
(811 to 7355 ohms).
Is RT1
temperature
between
32°F and 122°F
(0°C and 50°C)
?
Suspect RT1.
Repair/replace.
Is it
between
811 and 7355
ohms
?
Measure
resistance between
pins 2 and 9 of
connector P6.
Disconnect
connector P6 from
J2 on power supply
board.
RFU CODE 9
Reservoir probe
reading faulty
SERVICE MANUAL