Pump
Pump motor should be oiled once a year with 3-4 drops of general purpose
motor oil in the locations identified on the pump label.
Panel Exterior
Clean the control panel and panel exterior with a cloth dampened with
isopropyl alcohol.
Blankets
Outside surfaces of standard blankets may be cleaned with a damp cloth and mild
detergent to prevent algae growth. To clean the insides, attach the blankets to
the machine and follow instructions for section 7.2, p. 18,
CLEANING
PROCEDURES,
Fluid System.
NOTE: Exposure to harsh chemicals will cause blankets to lose flexibility and
resistance to cracking.
Probe Check Well
The Probe Check Well should be cleaned with a small tubular brush and
detergent, and then wiped with a commercial disinfectant.
When performing the PROBE CHECK, use a disposable protective
sheath (Becton-Dickinson catalog 3700 oral sheath or equivalent)
on the probe.
Failure to use sheath could result in cross-contamination.
Probes
Do not autoclave. Clean with a damp cloth and mild detergent. Wipe dry.
Probes are made of PVC. If probe must be sterilized, use any cold sterilization
means except alcohol, which may accelerate probe deterioration. Always wipe
dry.
If gas sterilization is necessary, probes should be handled like any other PVC
product. Probes are made of PVC.
Reusable probes may be cleaned with a damp cloth and mild detergent. Always
wipe dry. Exposure to harsh chemicals will cause probe to lose flexibility and
resistance to cracking.
Disposable probes should be discarded after use.
NOTE: Exposure to harsh chemicals will cause probe to lose flexibility and
resistance to cracking. Do not use damaged temperature probes.
Discard probes having visible pinholes, cracks or abrasions.
7.2
CLEANING
PROCEDURES
(continued)
WARNING
CLEANING PROCEDURES
19
SERVICE MANUAL
Medi-Therm II
Содержание MEDI-THERM II MTA5900 Series
Страница 76: ...SERVICE MANUAL Medi Therm II www gaymar com P N 10187 000 7 98 SERVICE INFORMATION Figure 13 Test setup 72...
Страница 82: ...SERVICE MANUAL Medi Therm II www gaymar com P N 10187 000 7 98 This page has been intentionally left blank 78...
Страница 83: ...MEDI THERMTM II IS MADE IN THE USA www gaymar com...