SERVICE MANUAL
Medi-Therm II
REPAIR PROCEDURES
w
w
w
.g
aymar
.com P/N 10187-000 7/98
4.
Insert patient probe assembly into mounting hole and fasten with hex nut.
Insure patient probe assembly is oriented so that the bevelled corner of the
probe assembly is positioned as shown in figure 16, p. 76.
5.
Connect the digital control panel plug P4 to the board at J4.
6.
Connect the patient probe assembly plug P2 to the board at J2.
7.
Perform a voltage check of the power supply board (see section 9.2, p.
60).
8.
Connect the 26 conductor cable assembly from the power supply board.
9.
Perform the
FUNCTIONAL CHECK AND SAFETY INSPECTION
(section 7.3).
To install a complete new cover:
1.
Unplug the power cord.
2.
Remove old top cover from machine by removing 6 screws underneath
top cover and disconnecting cables. See figure 14, p. 73.
3.
Install new top cover.
4.
Perform the
FUNCTIONAL CHECK AND SAFETY INSPECTION
(section 7.3).
To replace thermostat (see figures 15, 17, 17A and 18, pp. 74-75, 77, 79):
1.
Unplug the power cord.
2.
Remove wraparound from chassis.
3.
Remove thermostat cover. (On certain models only.)
4.
Disconnect spade lugs from thermostat terminals.
5.
Carefully peel back insulating material.
6.
Remove the two screws holding the thermostat. Remove the thermostat.
7.
Apply thermal grease * (P/N 50040-000) to the underside of the
replacement thermostat to ensure proper operating temperature ranges.
* Recommended thermal greases:
•
Silicone Heat Sink Compound
(Dow Corning; Midland, Michigan)
•
Thermal Joint Compound Type 120-5
(Wakefield Engineering; Wakefield, Massachusetts)
Always use thermal grease on thermostats for MTA5900
series Medi-Therm II units.
Failure to use thermal grease could raise trip points
above the acceptable range.
8.
Reconnect spade lugs to terminals. Refer to figures 17 & 17A, p. 77 to
ensure thermostat is connected correctly.
9.3
REPLACING THE
CONTROL/DISPLAY
BOARD
(continued)
9.4
REPLACING THE
TOP COVER
9.5
REPLACING
THERMOSTATS
CAUTION
62
Содержание MEDI-THERM II MTA5900 Series
Страница 76: ...SERVICE MANUAL Medi Therm II www gaymar com P N 10187 000 7 98 SERVICE INFORMATION Figure 13 Test setup 72...
Страница 82: ...SERVICE MANUAL Medi Therm II www gaymar com P N 10187 000 7 98 This page has been intentionally left blank 78...
Страница 83: ...MEDI THERMTM II IS MADE IN THE USA www gaymar com...