SERVICE MANUAL
Medi-Therm II
w
w
w
.g
aymar
.com P/N 10187-000 7/98
TROUBLESHOOTING CHARTS
No
Yes
Assume Q4 triac is
also damaged.
Repair/replace
power supply
board.
Replace
pump.
Suspect Q4 or U2.
Repair/replace
power supply
board.
Is
pump
winding
shorted
?
Yes
Yes
Yes
Yes
No
No
No
No
Suspect
control/display
board.
Repair/replace.
Suspect
board-to-board
connection.
Repair.
Is pin
8 ref to pin
10 of U51
0 VDC?
Disconnect pump and
measure resistance for
short. (It should be
approximately 10
ohms.)
Is
optocoupler U2
input pulled
low?
Suspect
base-to-power
supply board
connection.
Repair.
Line
voltage at
output of
Q4?
Repair/
replace
pump.
Line
voltage at
terminals 6
and 8 of
TB1?
Verify unit is in
AUTO or MANUAL
modes.
PUMP MOTOR
NOT RUNNING
Figure 6J—Pump Motor Not Running
50