SERVICE INFORMATION
Item
Name
Function
1
FLOW-OK
Indicates blanket water is flowing adequately.
2
HEAT
Indicates blanket water is being heated. Refer to section 4.2 THERMAL SPECIFICATIONS, p..5
3
IN-TEMP
Indicates the Medi-Therm II is controlling temperature at the desired setting.
4
COOL
Indicates blanket water is being cooled. Minimum blanket temperature is 4°C (39.2°F).
5
ALERT
Indicates an alert condition exists.
6
SELECT
Indicates a mode must be selected. Flashes if set point buttons are pressed before selecting mode.
7
CHECK WATER FLOW Indicates the flow of water to the blanket has been restricted.
8
CHECK PATIENT
PROBE
Indicates the patient probe is sensing an abnormal patient temperature (below 32°C [89.6°F] or above
45°C [113°F]). An audible alarm accompanies this alert.
WARNING: It may indicate the patient probe has become dislodged, an incorrect probe is being
used, or a probe has become defective.
9
ADD WATER
Indicates the water reservoir is below the recommended level.
10
REMOVE FROM USE
NOW / MACHINE
SHUT DOWN
Indicates the Medi-Therm II has shut down due to blanket water temperature exceeding either the high
or the low limit. It may also indicate the microprocessor has shut down the machine.
Remove the Medi-Therm II machine from use immediately.
11
TEST LIGHTS
Enables the operator to confirm that all indicator lights and digital displays are working, and that the
audible alarm is functioning. The displays flash and the audible alarm sounds as long as the TEST
LIGHTS button is pressed.
12
SILENCE ALARM
Temporarily silences the audible alarm The alarm will reactivate in approximately 5 minutes if the
condition persists. The SILENCE ALARM button will not function if the REMOVE FROM USE NOW /
MACHINE SHUT DOWN indicator is lit.
13
MANUAL
Selects the manual mode of operation. In MANUAL mode, the operator must observe patient
temperature and adjust the blanket water set point temperature to obtain the desired results.
14
AUTO
Selects the automatic mode of operation. In AUTO mode, the operator sets the desired temperature
and the patient temperature is automatically regulated to the operator-specified set point.
15
MONITOR
Selects the monitor mode of operation. In MONITOR mode, the operator may monitor patient
temperature without providing therapy.
16
BLANKET
Indicates the actual temperature of the water being delivered to the blanket. This display is illuminated
when MODE is set to either MANUAL or AUTO.
17
PATIENT
Indicates actual patient temperature as sensed by the probe connected to the patient probe jack. This
display is illuminated when MODE is set to MANUAL, AUTO, or MONITOR and a patient probe is
plugged into the PATIENT PROBE jack.
18 *
°C/°F INDICATOR
Displays the temperature and setpoint in either degrees Celsius (°C) or Fahrenheit (°F).
Pressing the °C/°F button will display the appropriate scale and light the corresponding indicator.
19 *
°C/°F BUTTON
20
LEADER LIGHTS
Emphasizes whether blanket temperature or patient temperature is being set.
21
22
SET POINT
SET POINT BUTTONS
Indicates the desired SET POINT temperature. When operating in MANUAL mode, pressing the SET
POINT button (up or down) sets the desired BLANKET water temperature. When in AUTO mode,
pressing the SET POINT button sets the desired patient temperature.
23
OPERATING
INSTRUCTIONS
Condensed operating instructions are mounted on the machine head and on top of the water fill
opening cover.
24
WATER FILL
OPENING
Used to fill water reservoir to proper level. To fill, lift cover on the water fill opening and fill with distilled
water until the green band on the float stem is fully visible.
CAUTION: Add distilled water only. Failure to use distilled water may result in poor machine
performance. Do not use alcohol. Do not operate without water. Do not overfill.
25
PATIENT PROBE
JACK
Accepts patient probe plug connecting patient probe to machine.
Use only YSI 400 series probe or equivalent.
26
BLANKET
CONNECTIONS
Connects machine to quick-disconnects on the connector hose.
Quick-disconnect fittings are marked BLANKET 1 and BLANKET 2.
27
PROBE CHECK WELL
Used to perform probe check (refer to Operating Manual for procedure).
WARNING: When performing the probe check, use a protective sheath (Becton-Dickinson
catalog 3700 oral sheath or equivalent) on probe to prevent cross-contamination.
28
ON / OFF SWITCH
(Circuit breaker) Controls power to the unit.
29
AIR VENTS
Permits adequate air circulation. Keep louvered openings clear on all three sides of the cabinet.
* Some models do not have the °C/°F feature
Table 6—Operator Controls/Indicators