Quant Model Name
Displays the quantification model name of
the actual substance calibration file.
INFORMATION!
All substance calibration information are part of the calibration file
set in the G.A.S. software VOCAL.
For detailed Information refer to the
VOCAL Tutorials.
INFORMATION!
The calibration file must be named as
calibration.gsd
6.5.5 Recognised substances window
When substances are enabled in the substance window the recognised substances
window appears at the end of every measurement. This screen displays a list with
the results in a concentration unit which is defined in the valid calibration curve.
When a current loop is available the last result of this list will be overwritten. By
starting the next measurement this window disappears until the next measurement
will be finished.
Figure 41: Recognized Substances Window