5.3 Principle setup and internal gasflow
The schematic drawing shows the principle structure of the gas flow system of the GC-
IMS is shown. The system consists of the IMS coupled to a gas chromatographic
column.
Figure 6: GC-IMS device plan (exemplary)
The drift gas for the IMS is supplied using an electronic pressure control unit
(EPC1). The carrier gas for the column is supplied using a second electronic
pressure control unit (EPC2). Both gases (carrier and drift gas) leave the device at
the gas out, which should be connected to an exhaust system.
INFORMATION!
To ensure correct measurements it is absolutely necessary to
connect the supplied exhaust tubes
(Gas out and Sample gas
out).
The exhaust tubes (Gas out and Sample gas out)
must be led
separately into the exhaust system and must not be
connected.
The exhaust system must
not generate any negative pressure
.
IMS (T1), GC-Column (T2), 6-Port Valve with sample loop (T3) and Transferline (T5
are heated.
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Страница 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...