GC-IMS
– User Manual
197/214
Action
Do a measurement. Repeat the procedure.
If
that still does not help contact the G.A.S.
service hotline.
Error message
RTC read error.
Description
Date and time setting failed.
Action
Restart the system and repeat the procedure.
If that still does not help contact the
G.A.S. service hotline.
Error message
Save settings failed.
Description
Setting data could not be save on a USB-stick.
Action
Check write protection of the USB-Stick.
Ensure that the USB-Stick is FAT32 formatted.
Reformat the USB-Stick. Replace the USB-Stick
with a new one. Repeat the procedure.
If that
still does not help contact the G.A.S.
service hotline.
Error message
Load settings failed.
Description
Setting data could not be load from a USB-stick
Action
Repeat the procedure. Restart the system and
repeat the procedure. Check write protection
of the USB-Stick. Ensure that the USB-Stick
is FAT32 formatted. Reformat the USB-Stick.
Replace the USB-Stick with a new one. Repeat
the procedure.
If that still does not help
contact the G.A.S. service hotline.
Error message
Do you really want to relabel loop volume? It will be
stored in all following measurements.
Description
Sample loop setting is change manually.
Action
Confirm or abort the dialog.
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Страница 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...