GC-IMS
– User Manual
90/214
6.7.5.4 Standby Mode Window
By activating the Standby Mode, the
Standby Mode Window
appears. The flowrate
of the Drift gas (EPC1) and the carrier gas (EPC2) will be decrease to reduce gas
consumption. The
standby mode flowrate
for drift gas
(EPC1) is 10 ml/min
and for
carrier gas
(EPC2) is 5 ml/min
.
Figure 61: Operating Interface - Standby Mode Window
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Страница 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...