
GC-IMS
– User Manual
142/214
3
In
Projection: User Value
-> Current
set
Min
and
Max
value of the
concentration range to the
range 0 mA
– 20 mA.
4
In
Calibration: Current ->
DAC
set
DAC [0 mA]
and
DAC [20 mA]
calibrate to
0-20 mA by observing the
outputted mA value on the
external display unit. The
values on the picture are
examples. The values for
DAC [0 mA] and DAC [20
mA] are preset by G.A.S.
but
must be adapted by
the user.
INFORMATION!
The values for DAC [0 mA] and DAC [20 mA] are preset. The correct
values for DAC [0 mA] and DAC [20 mA] must be adapted by the user.
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Страница 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...