Wait until cleaning
process is completed.
The process can be
stopped with
Skip
.
6
After all temperatures
reached the default
values inspect visually
the current spectrum and
compare it with the
reference spectrum of the
analytical approval.
The
RIP should reach ~80%
of the RIP height
displayed in the
delivered Analytical
Approval of the device
under same
measurement
conditions of G.A.S.
If not check the gas
quality and/or install
additional purification
cartridges and start the
cleaning procedure
again.
INFORMATION!
The device is delivered with an acceptance snapshot. This
snapshot defines the system performance during device
acceptance and is used to assess the readiness for measurement.
Any deviations from this are displayed in the Error Information
Window.
The default values can be adjusted by the customer
.