Action
Using the pushable rotary knob to active the
screen.
If the screen can’t be reactivated
then restart the system. If that still does
not help contact the G.A.S. service hotline.
Symtom
Gasflow set-values cannot be achieved.
Possible Cause
Backpressure to low
Action
Set up the back pressure at least to 3 bar.
If that still does not help contact the
G.A.S. service hotline.
Possible Cause
Hardware failure
Action
Contact the G.A.S. service hotline.
Symtom
Temperature set-values cannot be achieved.
Possible Cause
Hardware failure
Action
Contact the G.A.S. service hotline.
Symtom
No display of measurement values during monitoring
Possible Cause
Hardware failure
Action
Contact the G.A.S. service hotline.
Symtom
No Reaction Ion Peak (RIP) will be displayed.
Possible Cause
Device is in negative mode
Action
Switch the drift voltage into positive mode.
If that still does not help contact the
G.A.S. service hotline.
Possible Cause
Hardware failure
Action
Contact the G.A.S. service hotline.
Symtom
No or small Reaction Ion Peak (RIP) will be displayed.
Possible Cause
Insufficient gas quality
Action
Check quality of operating gas (5.0 or
better) and use moisture trap.
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Страница 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...