Fatal Error!
Contact the G.A.S. service
hotline
.
Error message
Invalid program.
Description
The selected program has no instructions.
Action
Complete the program and repeat the
procedure.
If that still does not help
contact the G.A.S. service hotline.
Error message
Not enough storage space left. Please delete or copy
measurement or choose a shorter program.
Description
The internal storage has not enough storage space left.
Action
Export and save all Measurement of the
internal storage and clear the storage.
Repeat the procedure.
If that still does not
help contact the G.A.S. service hotline.
Error message
Hardware access failure
Description
Error when switching the drift voltage polarity.
Action
Fatal Error!
Contact the G.A.S. service
hotline
.
Error message
Valve set to Inject. Loop not filled! Proceed? OK or
Chancel
Description
Valve start position is set to Inject instead of Fill loop.
Action
Set valve start position manually to Fill
loop. Repeat the procedure.
If that still
does not help contact the G.A.S. service
hotline.
Error message
Can’t import calibration.gsd
Description
The file „calibration.gsd“ was not found.
Action
Create the file „calibration.gsd“
and import
it with an USB-Stick. Repeat the procedure.
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Страница 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...