![GAS GC-IMS Скачать руководство пользователя страница 191](http://html1.mh-extra.com/html/gas/gc-ims/gc-ims_user-manual_3647276191.webp)
GC-IMS
– User Manual
191/214
6
Unused
Substance at Register n
Register
Datatype
Information
n
32bit unsigned Integer
The
concentration
(multiplied by 1000 for a
possible three decimal point
precision
) encoded into a
32bit Integer split up into this
and the following register.
The unit for this concentration
is specified in the
corresponding calibration and
can be seen in Substance
window of your device. The
precision of this calculated
value can be looked up in the
corresponding calibration
protocol for this substance.
Contains the two least
significant bits.
n+1
Second part of the encoded
concentration
(multiplied by
1000 for a possible three
decimal point precision) from
register n. Contains the two
most significant bits.
n+2
16bit unsigned Integer
Shows you whether the
substance has been
enabled
(=1)
or
disabled (=2)
for
evaluation.
n+3
16bit unsigned integer
A 16bit encoded status of the
substance result.
See
section Status codes
for
further information.
n+4
-
Unused
n+5
-
Unused
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...