Action
Restart the system. If that still does not
help contact the G.A.S. service hotline.
Symtom
Six-Port-valve switching noise sounds strange
Possible Cause
Initialization lost
Action
Reintializise the Six-Port-Valve
Symtom
No sample signals are displayed during a measurement.
Possible Cause
Sample-Pump has failed
Action
Set pump capacity to 100%. The pump device
must be audible loud. Turn the pumping power
back to the working flow. In addition, the
flow can be measured using a flowmeter at the
sample inlet and outlet. If that still does
not help change sample pump.
Symtom
Drift time fluctuates
Possible Cause
System is untight
Action
Check all gas supply connections of operating
gas.
Action
Check all gas supply connections of GC-
column.
Symtom
Retention time fluctuates
Possible Cause
System is untight
Action
Check all gas supply connections of operating
gas.
Action
Check all gas supply connections of GC-
column.
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Страница 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...