
xiii
EMC Guidance
This equipment complies with IEC 60601-1-2 (2007). However, if portable
transmitter or wireless LAN equipment is used extremely nearby, the
electromagnetic influence may largely exceed the compliance level and may
cause unexpected phenomenon such as noise interference on the waveform,
etc.
Therefore, this equipment should be used in a location specified by each
medical institution. If any unexpected noise interference on the waveform or
failure to the peripheral device occurs, stop using the equipment and follow the
instruction of the technician.
The following is the information relating to EMC (Electromagnetic
Compatibility).
(When using this equipment, verify that it is used within the environment
specified below.)
Compliance to the Electromagnetic Emissions
The LX-7230N is intended for use in the electromagnetic environment
specified below. The customer or the user of the LX-7230N should assure that
it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment –
Guidance
RF Emission
CISPR 11
Group 1
The LX-7230N uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF Emission
CISPR 11
Class A
This LX-7230N is suitable for use in all
establishments other than domestic
establishments.
Harmonic Emission
IEC 61000-3-2
N/A
Voltage Limit /
Flicker Emission
IEC 61000-3-3
N/A
Содержание LX-7230N
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 66: ...7 Measurement 46 Blank Page ...
Страница 72: ...8 Operation 52 Blank Page ...
Страница 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Страница 84: ...11 Troubleshooting 64 Blank Page ...
Страница 92: ...14 Standard and Optional Accessories 72 Blank Page ...