11. Troubleshooting
63
SpO
2
Phenomenon
Cause
Solution
SpO
2
value is
unstable.
The probe size is
improper.
Use a probe, which fit
properly.
The probe is
peeling off or is
affected by the
outside light due to
the poor condition
Attach the probe properly
following the instruction.
Transmitting and
measuring LEDs
sensor are dirty.
Clean both LED sensors
from dirt.
In Case of Dropping the LX-7230N into Water
In case of dropping the LX-7230N into water containing disinfectant, pick up
the LX-7230N quickly from it. Rinse it well with running water, and dry it
thoroughly with a soft cloth.
C A U T I O N
Do not use a drier. The LX-7230N shape may change or be broken.
When the LX-7230N is rinsed with running water, make sure to close the
battery compartment lid.