Applying the OxiMax
®
MAX-N sensor
This Nellcor
®
adhesive sensor, model MAX-N, is indicated for continuous
noninvasive arterial oxygen saturation and pulse rate monitoring and can be
reused on the same patient as long as the adhesive tape attaches without
slippage.
1. Remove the plastic backing from the MAX-N and locate the two
transparent windows on the adhesive side. Windows cover optical
components. Note the corresponding alignment marks (a) on the non-
adhesive side and the dashed line (b) midway between the marks (Figure
(1)).
2. Orient the MAX-N so that the dashed line is on the lateral edge of the site
(a):
Neonates:
The preferred site is the foot. Alternatively, use the hand. The
window next to the cable goes on the sole of the foot as shown (Figure (2)).
Adults:
The preferred site is the index finger. Alternatively, other fingers
may be used. The window next to the cable goes on the nail side, distal to
the first joint. Do not place on a joint. Note that the cable must be
positioned on the top of the hand (Figure (3)).
N O T E
When selecting a sensor site, priority should be given to an extremity free of
an arterial catheter, blood pressure cuff, or intravascular infusion line.
3. Wrap the MAX-N firmly, but not too tightly around the foot or finger.
Windows must oppose each other.
4. Connect the MAX-N into the LX-7230N. Verify proper operation as
described in the operation manual.
(1)
(2)
(3)
Содержание LX-7230N
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 66: ...7 Measurement 46 Blank Page ...
Страница 72: ...8 Operation 52 Blank Page ...
Страница 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Страница 84: ...11 Troubleshooting 64 Blank Page ...
Страница 92: ...14 Standard and Optional Accessories 72 Blank Page ...