12. Cleaning and Disinfection
66
Cleaning the ECG lead cable
After using the cable, clean it with neutral detergent or 70% isopropyl alcohol.
C A U T I O N
Do not use thinner, toluene, or other organic solvents to clean the cables.
Do not pull the cable and do not hold the connector part when cleaning.
(It may degrade the cable coating and result in damage. Particularly
organic solvents and antiseptic solution such as cresol soap solution will
degrade the cable coating.)
After cleaning, dry it completely before usage.
Do not use high temperature sterilization such as steam or EOG method.
Cleaning and Disinfecting the SpO
2
probe (sensor)
NELLCOR
®
SpO
2
sensors
After using the the Durasensor
®
(DS-100A) on each patient, clean it with 70%
isopropyl alcohol. If low-level disinfection is required, use a 1:10 bleach
solution. Any unproved cleaning solutions may case early failure of the
sensors.
The OxiMax
®
can be reused on the same patient as long as the adhesive tape
attaches without slippage. Do not resterilize and reuse it on other patients. It is
intended for single patient use only.