Precautions about Output Signal
W A R N I N G
Do not use the output signal of the monitor that receives radio wave signal
from the LX-7230N as the trigger signal for IABP, MRI echocardiographic,
or defibrillator for the following reasons.
It may lead to a delay of operating timing due to the delay time of
waveform transmission.
A trigger signal unrelated to the heart rate may be generated due to
the interfusion of spike noise at weak electric field.
Precautions about Accessories and Optional Accessories
W A R N I N G
Use only the accessories, such as ECG Lead cable and SpO
2
probe
(sensor), specified by Fukuda Denshi for the LX-7230N. Otherwise, the LX-
7230N cannot deliver its maximum performance and may be damaged,
resulting in a safety hazard.
C A U T I O N
Do not reuse disposable products.
Store the disposable products properly as mentioned in their user
manuals.
Precautions about Battery
W A R N I N G
Use new "AA" size (“LR06” size) alkaline cell.
Install the battery with the correct polarity.
Do not charge the battery. Any attempt to charge the battery may cause
it to leak or break.
Do not short the (+) and (-) terminals. It may result in exothermic heat
and fire.
Do not throw the battery into fire. It may explode.
Содержание LX-7230N
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 66: ...7 Measurement 46 Blank Page ...
Страница 72: ...8 Operation 52 Blank Page ...
Страница 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Страница 84: ...11 Troubleshooting 64 Blank Page ...
Страница 92: ...14 Standard and Optional Accessories 72 Blank Page ...