6. SpO
2
Monitoring
18
SpO
2
Monitoring
LX-7230N, which has a built-in Nellcor
®
SpO
2
module, is described in this
section.
The optional SpO
2
sensors available for LX-7230N are as follows.
The following table shows applicable patient and proper site for each SpO
2
sensor. Select the proper one depending on the purpose and intended use.
Sensor Types
Applicable Patient
Applied Site
OxiMax
®
MAX-I
Infant
(weight of 3 to 20Kg)
Toe
OxiMax
®
MAX-P
Pediatric
(weight of 10 to 50Kg)
Finger
OxiMax
®
MAX-A/AL
Adult
(weight of 30Kg and over)
Finger
Durasensor
®
DS-100A
Adult
(weight of 40Kg and over)
Finger
OxiMax
®
MAX-R
Adult
(weight of 50Kg and over)
Nose
OxiMax
®
MAX-FAST
Adult/Pediatric
(weight of 10Kg and over)
Forehead
OxiMax
®
MAX-N
Adult
(weight of 40Kg and over)
Finger
Neonate
(weight of less 3Kg)
Foot
W A R N I N G
For SpO
2
monitoring, always use the sensor specified by Fukuda
Denshi. If any other sensor is used, high temperature rise of the sensor
may place the patient in danger of burns in the worst case.
As with all medical equipment, carefully route cables to reduce the
possibility of patient entanglement and strangulation.
Содержание LX-7230N
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 66: ...7 Measurement 46 Blank Page ...
Страница 72: ...8 Operation 52 Blank Page ...
Страница 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Страница 84: ...11 Troubleshooting 64 Blank Page ...
Страница 92: ...14 Standard and Optional Accessories 72 Blank Page ...