11. Troubleshooting
60
SpO
2
Message
Cause
Solution
Probe is off.
Check the attached condition
of the probe.
Faulty Probe.
Replace the probe with a
new one.
SpO
2
is not
measured correctly.
Check the attached condition
of the probe.
Cover the probe with an
opaque material to cut off
the outside light.
Due to excessive
body motion, SpO
2
is not measured
correctly
Keep the patient still.