17
6. SpO
2
Monitoring
When the transmitter is used without the ECG lead cable, it will measure only
SpO
2
.
W A R N I N G
When the SpO
2
probe (sensor) is in a connector-off condition, the SpO
2
alarm will not be generated on the receiving monitor. Make sure that the
SpO
2
probe (sensor) is securely connected. If the SpO
2
waveform/numeric
data is not displayed, check the patient’s condition and pay attention not to
miss the connector-off condition.
C A U T I O N
When using the transmitter with only the SpO
2
sensor cable, ECG and
respiration measurements on the receiving monitor shall be turned off to
prevent an erroneous alarm.
The pulse wave and level meter are normalized for SpO
2
measurement.
It does not indicate perfused blood volume. Check proper probe
attachment by observing the pulse wave.