15. Specification
77
obtained from simultaneously sampled arterial blood using a laboratory
Co-oximeter.
Many functional testers and patient simulators have been designed to
interface with the pulse oximeter's expected calibration curves and may be
suitable for use with pulse oximeters and/or sensors. Not all such devices,
however, are adapted for use with the calibration system. While this will
not affect use of the simulator for verifying system functionality, displayed
SpO
2
measurement values may differ from the setting of the test device.
For a properly functioning pulse oximeter, this difference will be
reproducible over time and from pulse oximeter to pulse oximeter within
the performance specifications of the test device.
Displays
The following displays are shown on the LCD of the LX-7230N.
Display
Description
Starting Screen
Displays after turning on the power.
Automatically moves onto the channel
display screen.
Channel Display
Screen
Displays the transmitter channel after
turning on the power and also when
refreshing the screen.
Automatically move onto the waveform
display screen.
EVENT
Displays when the EVENT button is
pressed.
Automatically move onto the channel
display screen.
ELECTRODE?
Displays when the ECG electrode is
disconnected or the ECG/respiration
waveform cannot be measured normally.
For details about electrode check
message, refer to page 79.
Содержание LX-7230N
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 66: ...7 Measurement 46 Blank Page ...
Страница 72: ...8 Operation 52 Blank Page ...
Страница 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Страница 84: ...11 Troubleshooting 64 Blank Page ...
Страница 92: ...14 Standard and Optional Accessories 72 Blank Page ...