5. Secure the NIBP holder with the 4 screws.
6. Press the
(Menu) key
→
select
"Maintenance"
→
enter the password
→
select
"Periodic Replacement Parts Information", and
select [Reset] to reset the number of NIBP
measurement on the "Periodic Replacement P
Information" window.
arts
C A U T I O N
Check if the air hose is properly connected. Bad connection may cause
leakage of air and BP cannot be correctly measured.
7. After the power is turned ON, check that initialization of the NIBP unit is properly completed.