TCON System Construction (For USA Only)
Bidirectional communication to the central monitor is possible by using the Bidirectional Wireless
Communication Module (HTC-702).
y
Measured data can be sent to the central monitor.
y
Alarm settings can be synchronized.
y
NIBP can be measured from the central monitor.
TCON System Example
This section describes the TCON system example of the bedside monitor and the central monitor.
<Configuration Example of medical telemetry and TCON>
C A U T I O N
z
Check that three antenna bar marks (
) are displayed.
z
Make sure that the TCON channel number between the bedside monitor
and central monitor is the same.
z
Do not move the equipments during TCON operation. The radio waves
may not be transmitted.
z
The alarm limit of HR and PR is linked with each other. Changing the HR
alarm limit will also change the PR alarm limit to the same value. The
alarm ON/OFF setting of HR and PR is also linked with each other.
Changing the alarm ON/OFF setting for HR will also change the setting for
PR.
Bedside Monitor : DSL-8001 System
ID :
1
Channel :
9
Bedside Monitor : DSL-8001 System
ID :
16
Channel :
9
Central Monitor : DS-7700 System
ID :
1
Channel :
9
(Max. 16 bedside monitors)
7
-
11
Содержание DSL-8001
Страница 4: ...Blank Page ...
Страница 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 44: ...2 6 Blank Page ...
Страница 60: ...3 16 Blank Page ...
Страница 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Страница 62: ...Block Wiring Diagram This section describes the block wiring diagram of the DSL 8001 Block Wiring Diagram 4 2 ...
Страница 64: ...4 4 Blank Page ...
Страница 72: ...6 6 Blank Page ...
Страница 102: ...7 30 Blank Page ...
Страница 148: ...11 6 Blank Page ...
Страница 150: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100780 201301 ...