The <RESP Out of Range> message is displayed.
Cause
: The respiration rate is outside the measurement range.
Solution : Check if the electrodes are properly attached.
Check the lead cable.
The <RESP Suspended> message is displayed.
Cause
: RESP measurement is suspended.
Solution : Press the (Menu) key, and select “Parameter Setup” > “RESP Setup” > “RESP
Measurement”, and select [ON].
The <RESP Communication Error> message is displayed.
Cause
: Communication error with the ECG board has occurred.
Solution : Reboot the equipment. If the same message is repeatedly displayed, a failure of the
equipment can be considered. Cease the measurement and contact Fukuda Denshi.
Non-Invasive Blood Pressure
The monitor repeats the measurement, or “---” is displayed for the measurement value.
Cause 1 : The measurement accuracy is not reliable due to body motion artifact.
Solution : During the measurement, have the patient stay still as much as possible.
Cause 2 : The pulse is too small to acquire reliable measurement accuracy.
Solution : Check if the cuff is properly attached to the patient, or the cuff size is correct.
The time of measurement disappears and the measurement value is displayed as "---".
Cause
: The NIBP measurement data is deleted after the preprogrammed erase time
Solution : Select the appropriate time for "NIBP Erase Time" from 60min or 120min, which best
fits the monitoring purpose.
The cuff is not inflated although the pump is operating.
Cause 1 : The air hose is not firmly connected, and air is leaking.
Solution : Check if the air hose is properly connected.
Cause 2 : The cuff size does not match the selected patient type.
Solution : Check if the cuff size does match the selected patient type.
The <NIBP Check Cuff, Airhose> message is displayed.
Cause 1 : The connection between the cuff and air hose or the air hose and NIBP connector is
loose or disconnected.
Solution : If connection is loose or disconnected, securely connect it and perform the
measurement again.
If the same message is displayed again, air may internally leak. Cease the
measurement and contact our service representative.
Cause 2 : The cuff is compressed.
Solution : Make sure that the cuff is not subjected to compression, and measure with the patient
arms extended as much as possible.
If the same message is displayed again, a blockage in the air system can be
considered. Cease the measurement and contact our service representative.
The <NIBP Motion Artifact> message is displayed, and measurement has stopped.
Cause
: The measurement accuracy is not reliable due to body motion artifact.
Solution : During the measurement, have the patient stay still as much as possible.
8
-
5
Содержание DSL-8001
Страница 4: ...Blank Page ...
Страница 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 44: ...2 6 Blank Page ...
Страница 60: ...3 16 Blank Page ...
Страница 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Страница 62: ...Block Wiring Diagram This section describes the block wiring diagram of the DSL 8001 Block Wiring Diagram 4 2 ...
Страница 64: ...4 4 Blank Page ...
Страница 72: ...6 6 Blank Page ...
Страница 102: ...7 30 Blank Page ...
Страница 148: ...11 6 Blank Page ...
Страница 150: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100780 201301 ...