Accessories and Optional Accessories
W A R N I N G
Use only the cables specified by Fukuda Denshi.
Use of other cables may result in an increase in emission (generation
of unwanted electromagnetic energy from this equipment) or decrease
in immunity (resistance to electromagnetic interference generated
from other devices).
Precautions about the DSL-8001
D A N G E R
When connecting other equipment, contact your nearest representative.
Danger such as electric shock may result to the patient and operator.
W A R N I N G
z
Do not connect any equipment or cable not authorized by Fukuda
Denshi to any I/O connector. If done so by mistake, the device cannot
deliver its maximum performance and the connected units may be
damaged, resulting in a safety hazard.
z
If the equipment is used in an environment that does not fulfill the
specified conditions, the equipment cannot deliver the maximum
performance, equipment may become damaged and safety cannot be
ensured. If using in an environment other than the specified
environment, contact our service representative.
z
Use only the accompanying 3-way AC power cable. Use of other
cables may result in electric shock to the patient and the operator.
z
The power cable must be connected to a hospital grade outlet.
z
When using multiple ME equipment simultaneously, perform
equipotential grounding to prevent potential difference between the
equipments. Even a small potential difference may result in electric
shock to the patient and the operator.
z
Alarm Setup
y
Set the appropriate upper and lower alarm limit for each parameter
according to the monitoring condition.
y
When the upper/lower alarm limit of the individual parameter is set to
OFF
(
”- - -“
)
, alarm will not be generated even if the individual
parameter alarm is set to ON. Be cautious when setting them OFF.
y
HR/PR alarm and HR zero alarm is independent and the setting is
not linked with each other.
y
RESP alarm and RR zero alarm is independent and the setting is not
linked with each other.
z
Rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest. Do not rely entirely upon rate meter
alarms. Keep pacemaker patients under close surveillance. See this
manual for disclosure of the pacemaker pulse rejection capability of this
equipment.
z
The purpose of the RR zero alarm is to alert the user to evaluate for
the possible occurrence of apnea events by identifying the absence of
respiration. It is not intended to be classified as an “Apnea Monitor” and
will not identify the condition creating the possible event. (Central.
Obstructive or Mixed)
xiv
Содержание DSL-8001
Страница 4: ...Blank Page ...
Страница 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 44: ...2 6 Blank Page ...
Страница 60: ...3 16 Blank Page ...
Страница 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Страница 62: ...Block Wiring Diagram This section describes the block wiring diagram of the DSL 8001 Block Wiring Diagram 4 2 ...
Страница 64: ...4 4 Blank Page ...
Страница 72: ...6 6 Blank Page ...
Страница 102: ...7 30 Blank Page ...
Страница 148: ...11 6 Blank Page ...
Страница 150: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100780 201301 ...