7.2 Troubleshooting
If troubleshooting, check the fault by using the Hand Held Communicator (HHC) with function
"SELF CHECK" - Refer the chapter " Operation" in this manual.
The transmitter is supplied with the burnout adjusted : “fixed” output signal in case of failure.
If an abnormality occurrs in the process or transmitter, action should be taken with
references to the table below:
Symptom
Cause
Remedy
Output current
overshoots
scale
(The value is
more than the
upper limit of
the saturation
current.).
(1) The manifold valve does not open/
close normally.
(2) Pressure leak is occurring.
(3) Process piping is improper.
(4) Process pipe is clogged.
(5) Power supply voltage and/or load
resistance is improper.
(6) Voltage between the external
FRQQHFWLRQWHUPLQDOVRIDPSOL¿HU
unit is wrong.
=HURDQGVSDQRU¿[HGRXWSXW
current(4, 20mA) are not adjusted.
(8) Electronic unit is faulty.
Repair the valve so that it opens/closes normally.
Repair leak.
Make correct piping.
Eliminate the cause of clogging.
Make arrangement to obtain proper values.
Check for faulty cable, insulation, etc. and repair
as needed.
Readjust according to chapter 6.
5HSODFHWKHDPSOL¿HUXQLWDFFRUGLQJWRFKDSWHU
No output
current (The
value is less
than the lower
limit of the
saturation
current).
(1) Same as (1) to (4) above
(2) Power supply polarity is wrong.
(3) Power supply voltage is improper.
(4) Voltage between the external
connection terminals is wrong.
=HURDQGVSDQRU¿[HGRXWSXW
current(4, 20mA) are not adjusted.
$PSOL¿HUXQLWLVIDXOW\
Correct wiring according to chapter 4.1.
Make arrangement to obtain proper values.
Check for faulty cable, insulation, etc. and repair
as needed.
Read just according to chapter 6.
5HSODFHWKHDPSOL¿HUXQLWDFFRUGLQJWRFKDSWHU
Output current
error.
(1) Process piping is improper.
(2) Gas or solution is mixed in.
(3) Liquid density changes.
(4) Ambient temperature changes
widely.
=HURVSDQRU¿[HGRXWSXWFXUUHQW
(4, 20mA) has deviated.
$PSOL¿HUXQLWLVIDXOW\
Correct the piping.
Vent or drain the transmitter.
Perform density compensation.
Minimize the temperature change.
Readjust according to chapter 6.
5HSODFHWKHDPSOL¿HUXQLWDFFRUGLQJWRFKDSWHU
When the
indicator is
abnormal.
(1) An error display is appeared.
Refer to “Contents of message”