A-2 Span (URV) calibration
• Press the key on the screen
h
to select the span
calibration mode.
The measured value and unit on the screen (
i
) are
WKHVDPHDVWKRVHLQWKHQRUPDOPRGHDQG³ĸ´DQG
“SPAN” light up.
• Apply the actual input pressure on the screen
i
. Af-
ter checking the measured value, press the key.
• “SPAN” blinks on the screen
j
. Press the key on
the screen
j
to perform a span calibration at the in-
put pressure at the time.
To perform a span calibration at a point other than
100%, input an appropriate set value (%) (
k
) with
the and keys, and press the key.
3/ 8SSHUOLPLWRIDGMXVWPHQWSRLQWî
Setting range
• Select whether the span calibration value setting is
saved on the screen
l
.
Press the key to save the span calibration value
setting and return to the screen
i
.
Press the or key to cancel the setting and return
to the screen
i
.
• Check that the span calibration was performed as in-
tended.
Press the key to perform a span calibration again.
Press the or key to move to the next screen for
item name selection.