7.4 LIST OF ERROR CODES FROM CONTROL SOFTWARE
ERRO NO./DISPLAY
CAUSE
SOLUTION/CHECK POINTS
1. Heater
Heater not powered
Check to see if the cable from the pyrolyzer is
properly connected to the POWER connector
located on the rear of controller.
2. Sensor
Abnormal input from sensor
[Furnace or Interface]
Make sure the cable from the pyrolyzer is
properly connected to the SENSOR connector
located on the rear of the controller.
[Cryo-Trap]
Ensure that the temperature sensor connector
is short-circuited when MicroJet Cryo-Trap is
not used.
3. Overheating
Overheated.
Make sure the cable from the pyrolyzer is
properly connected to the SENSOR connector
located on rear of the controller.
If this does not resolve the problem, it may be
heater failure (Furnace or Interface). Or it may
be a cooling fan failure (Sample cup stand by
position). Contact your nearest sales office.
4. COM Port
USB cable contact failure
Re-seat the USB cable at the PC
11. Connection
Connection failure to
Auto-Shot Sampler
(1) Ensure that the cable from Auto-Shot
Sampler is properly connected to RS-232C
connector on the rear of the controller.
(2) Make sure the Auto-Shot Sampler is
powered on.
12. Cup chute
Cup chute not attached to the
Auto-Shot Sampler.
Properly attach cup chute to Auto-Shot
Sampler.
13. Sample cup retrieval
Sample cups not properly
retrieved.
(1) Make sure supply pressure of gas reservoir
is 300 kPa
– 500 kPa.
(2) Ensure that the valve control cable from the
gas reservoir is properly connected to the rear
of the Auto-Shot Sampler.
(3) Make sure sample cups are not deformed.
21.Cryo-trap Temperature
MJT sensor not properly
cooled.
(1) While the thermal exchange coil is at room
temperature, make sure the pressure is 250
–
300 kPa when Cooling is turned ON.
(2) Make sure there is ample amount of liquid
nitrogen.