4.4.2.1 Temperature calibration of pyrolyzer furnace
The temperature calibration of the pyrolyzer furnace is done by entering the actual temperature
measured by other means. Because the pyrolyzer furnace temperature greatly influences analytical
results, you should familiarize yourself with this operation and verify the calibration on a periodic
basis.
Ensure that the pyrolyzer furnace temperature is stabilized at the set temperature. The set
temperature is shown in A1, but this is automatically entered with the temperature defined in the
Single-Shot analysis mode.
(1) Enter the actual temperature measured in A2, using a reliable thermometer of your choice.
(2) Click the
“Save” button to finish the temperature calibration.
(3) Make sure that the temperature displayed matches the actual temperature measured as Fig.4.23.
If not, repeat the operations (1) through (3) described above.
(4) The interface temperature calibration can be done in a similar manner by entering the actual
value measured in B.
C
B
Fig. 4.22 Screen shot of temperature calibration of pyrolyzer and other set-ups
D
A2
E
F1
G
H
I
A1
F2
When performing a temperature calibration, follow the instructions in
the manual to measure the correct temperature. If wrong values are
entered, the control of temperature is not properly done.
CAUTION
!