4-20
Ver.1.36
4.4.3 Printing set-ups
Set-ups saved in the currently loaded file can be printed. Go to the menu bar, select
“File” – “Print”.
4.4.4 Set-up the Py Conditioning Table
From the menu bar, choose
“Tools” – “Py Conditioning Table”. The Py conditioning table shown in Fig.
4.24 is displayed. This allows you to perform the conditioning of the pyrolyzer furnace. Enter desired
date and time, pyrolyzer furnace temperature, and interface temperature, and click
the “Run” button
to activate the timer. In the example shown in Fig. 4.24, at 21:00 on March 10, 2011 the pyrolyzer
furnace temperature will be set to 700°C and the interface temperature will be 300°C. At 0:00 on
March 11, 2011 the pyrolyzer furnace temperature will be set to 100°C and the interface will be set to
200°C. Note that while this is running, no other methods can be changed.
Fig. 4.24 Set-up the Py Conditioning Table