Ver.1.36
8-1
CHAPTER 8 GUARANTEE OF BASIC PERFORMANCE
(REPRODUCIBILITY)
In a GC pyrolysis system, the path from the pyrolyzer to the detector via the separation column is very
important. Every element in the system (i.e., the pyrolyzer, interface, GC injection port, separation
column, detector crossover and detector) plays an important role in the reproducibility of the data.
Reproducibility (expressed as the %RSD of three sequential analyses) of the polystyrene EGA
thermogram and the polystyrene pyrogram are guaranteed to meet the following specifications:
Evolved Gas Analysis: Frontier Laboratories guarantees that the reproducibility of the apex of the
polystyrene peak is
≤0.3% RSD.
Pyrolysis: Frontier Laboratories guarantees that the reproducibility of the peak area ratio between
methyl stearate, added as internal standard, and the styrene trimer (SSS) is
≤2% RSD.
The analytical conditions and typical chromatograms are shown in Figures 8.1 and 8.2.
There is a risk of burn injury.
Since the pyrolyzer furnace can be very hot, ensure that the protective housing
cover is properly placed on the pyrolyzer.
CAUTION
HOT
When using solvents, wear protective gears such as safety goggles
and work only in a well-ventilated area free of ignition sources.
The ends of separation column and eco stick are very sharp. When
handling these items, wear safety goggles for your safety.
CAUTION
!
CAUTION
!