At the moment when the factor of the manufacturer is confirmed, the
previous user factor is deleted from the memory.
9.4.1 Calibration during operation
During this procedure, a fluid is delivered in a measuring container under realistic operation conditions
(flow rate, viscosity), which must be strictly adhered to.
The following points must be observed for a correct calibration of the counter:
■
■
Completely bleed the air from the system before the calibration is carried out.
■
■
Use a calibrated container which has a capacity of at least 5 liters and features exact measuring marks.
■
■
For calibration, carry out the discharge procedure at a constant flow rate corresponding to the rate
issued during normal operation. Fill the container until it is full.