18
GB
Operating instructions - mobiSERVE
■
■
Do not reduce the flow rate when the measuring scale of the container is almost reached during the
final period of discharge. At the end of the filling operation, the right procedure is to carry out short
refilling cycles at the normal operating flow rate.
■
■
After the discharge operation, wait a few minutes in order to ensure that all air bubbles have been
removed from the container. The correct value can be read only upon completion of this phase,
because at an earlier stage, the level may still drop in the container.
■
■
Carry out the procedure correctly as described in the folowing.
9.4.2 Calibration procedure during operation
Operation
Display-Configuration
1
NONE
Counter in normal mode, not counting
2
LONG PRESS ON THE TOTAL KEY
The mobiSERVE enters into the calibration mode, shows
the message TOTAL and the calibration factor used instead
of the total amount. The messages FACT and USER indicate
which one of the two factors (manufacturer or user ) is
currently being used.
3
LONG PRESS ON THE RESET KEY
The mobiSERVE indicates the TOTAL screen and the
resettable total amount indicator is zero. The counter is
ready to carry out a calibration during operation.
4
DISCHARGE IN THE CALIBRATED CONTAINER
Start the dispense in the calibrated container without
pressing a key.