19
Operating instructions - mobiSERVE
GB
The discharge procedure can be interrupted and started
at any time. Carry out the discharge until the fluid has
achieved the measuring scale of the calibrated container. It
is not necessary to achieve a certain amount.
Set value
Actual value
5
SHORT PRESS ON THE RESET KEY
The mobiSERVE is informed that the calibration discharge is
finished. Check to ensure that the discharge is completely
finished before this happens.
To calibrate the mobiSERVE, the value indicated by the
partial amount counter (example 9.800) must be brought
to the actual value indicated by the calibrated container.
On the lower left corner of the display, an arrow appears,
pointing upwards or downwards, thus indicating the
direction in which the USER K FACTOR value is changed.
Increase or reduce when the operations 6 or 7 are carried
out.
6
SHORT PRESS ON THE RESET KEY
Change of arrow direction. This procedure can be repeated
as often as required.
7
SHORT/LONG PRESS ON THE TOTAL KEY
The value indicated is changed in the direction defined by
the arrow:
■
■
One unit for each short press on the TOTAL key
■
■
Continuously, when the TOTAL key is kept depressed -
the first 5 units change slowly, then quickly.
■
■
If the value desired is exceeded, repeat the operation
from point 6 (6).