PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
82
Table 16.
SETUP File Choices, Factory Preferred Choice and Normal Value
VARIABLE
SOURCE
CHOICES
FACTORY
DEFAULT CHOICE
NORMAL VALUE
Atmospheric pressure
1. Internal
2. Normal
3. User
4. COM2
1. Internal
101.325 kPa
(14.6959 psi)
Ambient temperature
1. Internal
2. Normal
3. User
1. Internal
20 °C
Ambient relative humidity
1.Internal
2. Normal
3. User
1. Internal
45 %RH
Piston-cylinder
temperature
1. Internal
2. Normal
3. User
1. Internal
20 °C
Gravity
1. Local
2. Normal
3. User
1. Local
9.806650 m/s
2
Vacuum
1. COM2
2. Normal
3. User
1. COM2
0 Pa
Ready/Not Ready
Piston position
1. User
1. User
±
2.5 mm from
midfloat position
(default)
Ready/Not Ready
Maximum vacuum
reference pressure in
absolute by vacuum mode
1. User
1. User
5 Pa (default)
Note
The factory default SETUP choice is the SETUP choice setting for all variables in a new
SETUP file. SETUP file #1 is fixed to factory SETUP choices and cannot be altered.
OPERATION
To access the SETUP menu, press
[SETUP]
from the main run
screen. The display is:
Current SETUP: #01
1select 2view 3edit
“#01” in the upper right hand corner displays the number of the SETUP file that is currently active.
Select
<1select>
to select a different active SETUP file. Select
<2view>
to view the variable choices and
values of the active SETUP file. Select
<3edit>
to create or edit a SETUP file.
See proceeding sections 3.11.1 to 3.11.3 for detailed information on the operation of each SETUP function.
3.11.1 <1select>
PURPOSE
To change the active SETUP file number.
Caution
Changing the current SETUP file will change the source and in some cases
the values of the variables used by PG9000 in calculating defined
pressure. SETUP files should be selected and changed only by qualified
personnel. Incorrect SETUP file selection may cause out of tolerance
calculations and measurements.
Содержание PG9000 Series
Страница 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Страница 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Страница 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Страница 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Страница 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...