5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION
Page 159
© 2011 Fluke Calibration
The values of resistances at 0
°
C and the slopes, if desired, can be edited to
reflect the new calibration by pressing
[SPECIAL], [7cal]
and selecting
<4PCT1>
for the first PRT, or
<5PCT2>
for the second PRT.
OPERATION
To view the output of the piston-cylinder module temperature sensors, press
[SPECIAL], <7cal>, <4PCT1>
,
<1view>
or
[SPECIAL], <7cal>, <4PCT2>
,
<1view>.
The display is:
1.
Current reading of the selected piston-
cylinder module temperature sensor in
degrees Centigrade.
2.
Indication that this is a piston-cylinder
module temperature display, and which
sensor.
3.
Current value of the resistance slope relative
to temperature.
4.
Current value of the offset at 0 °C in ohm.
19.951 °C PCT1
RZ100.003
Ω S0.389
To adjust the values of
RZ
and/or
S
, press
[SPECIAL] , <7cal>
and select either
<4PCT1>
,
<2cal>
or
<4PCT2>
,
<2cal>
to access a screen in which the values of
RZ
and
S
can be edited. From here, press
[ESC]
and select
<1view>
to view the
individual PRT reading with the edited calibration coefficients applied.
To remove the mounting post PRTs:
Mounting Post PRT Removal
Remove the piston-cylinder module from the mounting post and replace with
the ORANGE storage plug.
Disconnect the PG9000 Platform from the PG9000 Terminal.
Turn upside down the PG9000 Platform so that the bottom of the platform is
up. Support the platform so that it does not tip over. A simple solution is to
place the platform upside down on a sturdy box which is smaller than the
outside platform dimensions and is also tall enough to allow the mounting
post to be suspended.
Remove the two socket head screws (2.5 mm Allen tool), securing the small
circular AMH DRIVE connector bulkhead plate to the platform rear panel.
PRT1
PRT2
Socket Head Screws
Figure 22.
PG9602 bottom panel detail.
Содержание PG9000 Series
Страница 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Страница 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Страница 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Страница 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Страница 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...