PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
106
OPERATION
Note
PG9000 is delivered with NO active password so access to the
User Level menu is open. The User Level is set to <2Low>.
User Levels can be changed freely until a password has been
created. Reset functions (see Section 3.12.9) do not affect the
password setting.
To access the User Level function, press
[SPECIAL]
and select
<2level>
.
If no password yet exists or if the correct
password has been entered, the display is:
1change user level
2edit password
Selecting
<1change user level>
brings
up the restriction menu:
Restriction: 1none
2low 3medium 4high
You can then select the current restriction level, or press
[ESC]
to return to the
main run screen.
Selecting
<2edit password>
displays the
user password and allows it to be edited.
Passwords can be up to six numbers in
length and cannot start with a zero.
Password: pppppp
0 disables password
If
0
is entered as the password value, then the password is made inactive and a
password will NOT be required to access the User Level menu. This is the
factory default with a security level of
<2low>
.
Caution
Once a password has been entered, the User Level cannot be
changed without reentering the password.
If there is an active password, the
PG9000 password entry screen appears.
PG9000 NS nnn-xx
Password: pppppp
The user must enter the user defined password or the factory secondary
password to proceed. When a password is entered correctly, operation proceeds
to the
<1change user level 2edit password>
screen.
The first field, <
nnnn>,
is the serial number of PG9000, followed by a second
field, <
xx>.
That counts the number of times that a secondary password has
been used. This second field increments each time a secondary password
is used. The third field, <
pppppp
>, is for standard password entry.
The factory
secondary
password is available in case the user’s password has
been misplaced or forgotten. It can be obtained by contacting a Fluke Calibration
Authorized Service Provider. The factory
secondary
password is different for all
PG9000’s and changes each time it is used.
3.12.5 <5remote>
PURPOSE
To configure the PG9000 COM1, COM2, COM3, COM4 and IEEE-488 communication ports.
To test COM1, COM2, COM3 and COM4 communications.
Содержание PG9000 Series
Страница 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Страница 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Страница 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Страница 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Страница 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...